Liquid Particle Counter Specifications
- Sample input: Low pressure: 0 to 6 bar – batch and online mode
- Volumetric particle sensor PAMAS HCB-LD-50/50: Maximum particle concentration: 24,000 particles per ml at a coincidence rate of 7.8 %
- Nominal flow rate: 25 ml/min
- Calibration according to common cleanliness standards and based on international calibration standards which are traceable to the NIST (National Institute of Standards and Technology).
-
Up to 32 free adjustable size channels may be selected and stored in the system. 3 different channel settings for polyglycol measurements are pre-set in the system:
- SAE AS 4059, channels A – F, 4 – 70 µm(c)
- Calibration in six size channels: > 4, > 6, > 14, > 21, > 38, > 70 µm(c)
- SAE AS 4059, channels B – F, 6 – 70 µm(c):
- Calibration in five size channels: > 6, > 14, > 21, > 38, > 70 µm(c)
- NAS 1638, 5 – 100µm:
- Calibration in the size channels: > 5, > 15, > 25, > 50, > 100 µm